LEX50
LEX50 Ex ta/tb certified resistance thermometer for zone 1, 2 and 20, 21, 22
Specifications
ATEX execution | II 1 D Ex ta IIIC T85°C...T300°C Da |
II 1/2 D Ex ta/tb IIIC T85°C...T300°C Da/Db | |
II 2 D Ex tb IIIC T85°C...T300°C Db | |
ATEX certificate | DEKRA 17ATEX0123 X |
IECEx execution | Ex ta IIIC T85°C...T300°C Da |
Ex ta/tb IIIC T85°C...T300°C Da/Db | |
Ex tb IIIC T85°C...T300°C Db | |
IECEx certificate | IECEx DEK 17.0046X |
Suitable for zone | 1, 2 and 20, 21, 22 |
Ingress protection | IP65 / IEC529 |
Ambient temp. range | -40°C...+75°C, this can be limited depending on the materials applied such as cables, connection heads, junction boxes and transmitters. |
Model | CE (showed), other Ex ta/tb certified heads possible |
Material | Aluminium, plastic or stainless steel |
Cable Entry | M20x1.5 standard, other threads possible |
Construction | Springloaded and replaceable |
Springaction | >10mm standard, >40mm optional |
Diameter | 6.0mm standard, other diameters possible |
Type | Mineral insulated (MgO) |
Execution | Single or Duplex Pt100 |
Connection | 3-wire system, 4-wire system optional |
Accuracy | According IEC 60751 Cl. B, A or AA |
Measuring range | Standard -20°C...+400°C, optional -220°C...+600°C |
Type | Nipple - union - nipple standard, other constructions possible |
Material | Stainless Steel |
Style / Connection | 1/2"NPT standard, other threads possible |
Length | 150 mm standard, other lengths possible |
Customisation possibilities
You can remark the customisation in your quote
Productcode | LEX50 |
---|---|
Sensor S or D | … |
Connection 3- or 4-wire | … |
Class B, A or AA | … |
Diameter "D" in mm | … |
Length "U" in mm | … |
Length "X" in mm | … |